Calibration of a pH meter with its probe

All pH meters need to be calibrated periodically. To know how often you should calibrate your pH meter, click here.


As all pH meters have different features (automatic calibration, manual calibration, auto buffer recognition, etc.), the best calibration explanation of your Hanna meter is in the instruction manual. If you no longer have the instruction manual, you can download a copy by clicking here.


General procedure

If your pH electrode is new or has not been used for a while, proceed with conditioning of your pH electrode prior to use (store overnight into HI 70300L storage solution).

Dip your pH electrode into HI 7007L pH 7.01 buffer first. Wait for the reading to stabilize and adjust the pH meter to read 7.01 on the display.

Rinse the pH electrode with distilled or tap water.

Dip the pH electrode into HI 7004L pH 4.01 or HI 7010L pH 10.01 buffer and wait for the reading to stabilize. Then adjust to read accordingly. Calibration is done and your meter is ready to use.

Note: Always start calibrations at pH 7 first and 4 or 10 second. The choice of the second buffer depends on your application: if you measure values below pH 7, use pH 4 as a second buffer. If you measure values above pH 7, use pH 10 as your second buffer. Some of our pH meter use 3 points calibration, keep in mind to always start with pH 7 first and others after.




Verification of a pH electrode

First look for scratches on sensor (glass bulb). If the sensor seems to be OK, make sure your pH sensor is well conditioned (soak sensor into HI 70300L storage solution overnight).

When conditioning is completed, dip your electrode into pH 7 and wait for the reading to stabilize (If it does not stabilize, perform a cleaning procedure). Adjust the reading to read 7.0 pH. If you can not reach the value (reading higher than 8.2 or lower than 5.8) do a cleaning procedure. When able to adjust to 7.0, switch to mV readings on your pH meter. You should have a reading of 0.0mV, ± 30mV. If the reading is higher than ± 50mV, you need to replace the electrode. If reading is below 50mV, proceed to rinse electrode with distilled or tap water and dip into pH 4 or pH 10 buffer. Switch in pH measurement mode and wait for the reading to stabilize. Adjust the reading to read 4.0 pH. If you can not reach the value (reading higher than 5.0 or lower than 3.0) do a cleaning procedure. When able to adjust to 4.0, switch to mV readings on your pH meter. You should have a reading of 180mV, ± 30mV. If the reading is higher than ± 50mV, you need to replace the electrode.




Cleaning & storage of a pH electrode

Note: By cleaning the electrode with the help of the proper cleaning solution, you insure optimum results and longer life to your valuable instrument.


Product Application
HI 7061 Cleaning Solution for General Use

This solution can be used for about any applications. It is not aggressive, so you should use it regularly and your electrode will be ready whenever you need to use it.

Cleaning time: 30 minutes to 1 ½ hour

Frequency: Once or twice a month or more depending on application
HI 7073 Cleaning Solution for Protein

This cleaning solution will stop the problem of having junction clogged with protein. The only thing you have to do is dip the tip of the electrode for 15 minutes, each time the response time begins to slow down.

Use 1g. of the powder for each 100 ml of solution.
HI 7074 Cleaning Solution for Inorganic Substances

This solution is especially made for inorganic substances but will be very effective in other applications as well. It is usually used in industrial applications. Probably the fastest cleaning solution you can use.

Cleaning time: 5 to 15 minutes

Frequency: Once or twice a month or more depending on application
HI 7077 Cleaning Solution for Oils

An oily or greasy substance makes usually a difficult task for taking pH from sample to sample without contaminating the next one (because oils and greases are difficult to rinse with water). By rinsing the electrode with the HI 7077 and with distilled water after each measurement, you will be ready for the next sample

Storage of pH electrode

AFTER USE: Always clean the pH electrode with plenty of clean water and store it in the storage solution (fill the cap with storage solution and put it back on). Verify the level regularly, as the solution will evaporate. Whenever you need to take measurements, the pH electrode will be ready to use as it will already be conditioned.



How to use replenishable junction

Over time, the junction which is the most sensitive part of the pH electrode can become clogged. This results in the electrode response becoming increasingly sluggish and eventually impossible to calibrate. With the new Hanna replenishable junctions, by using an ordinary pair of tweezers, simply pull out 1-2mm (1/8’’) of the fiber junction and you will literally have a reconditioned pH electrode. This procedure can be repeated up to 15 times, before the whole fiber gets out.

                          



ProductsSupportLiteratureInformationCompany

Bienvenue